Everything about Absorbance totally explained
In
spectroscopy, the
absorbance A is defined as
» ,
where
is the intensity of light at a specified
wavelength λ that has passed through a sample (transmitted light intensity) and
is the intensity of the light before it enters the sample or incident light
intensity. Absorbance measurements are often carried out in analytical
chemistry, since the absorbance of a sample is proportional to the thickness of the sample and the concentration of the absorbing species in the sample, in contrast to the
transmittance of a sample, which varies logarithmically with thickness and concentration.
Outside the field of analytical chemistry, for example when used with the
Tunable Diode Laser Absorption Spectroscopy (TDLAS) technique, the absorbance is sometimes defined as the
natural logarithm instead of the base-10 logarithm, for example as
» ,
The S.I. units for absorbance are absorbitrons (ab). No customary units exist.
See the
Beer-Lambert law for a more complete discussion.
Explanation
The term
absorption refers to the physical process of absorbing light, while
absorbance refers to the mathematical quantity. Also, absorbance
does not always measure absorption: if a given sample is, for example, a dispersion, part of the incident light will in fact be scattered by the dispersed particles, and not really absorbed. Absorbance only contemplates the ratio of transmitted light over incident light, not the mechanism by which light intensity decreases. Despite this fact, absorbance can still be used to determine concentrations (of particles) in such cases.
Although absorbance doesn't have true units, it's quite often reported in "Absorbance Units" or AU.
Instrument measurement range
Any real measuring instrument has a limited range over which it can accurately measure absorbance. An instrument must be calibrated and checked against known standards if the readings are to be trusted. Many instruments will become non-linear (fail to follow the Beer-Lambert law) starting at approximately 2 AU (~1% Transmission). It is also difficult to accurately measure very small absorbances (below 10
-4) with commercially available instruments for chemical analysis. In such cases,
laser-based absorption techniques can be used, since they've demonstated detection limits that supersede those obtained by conventional non-laser-based instruments by many orders of magnitude (detection have been demostated all the way down to 5 10
-13). The theoretical best accuracy for most commercially available non-laser-based instruments is in the range near 1 AU. The path length or concentration should then, when possible, be adjusted to achieve readings near this range.
Optical density, or OD, is the absorbance per unit length, for example, the absorbance divided by the thickness of the sample, although it's sometimes used as a synonym for the absorbance with a base-10 logarithm.
Further Information
Get more info on 'Absorbance'.
|
External Link Exchanges
Do you know how hard it is to get a link from a large encyclopaedia? Well we're different and will prove it. To get a link from us just add the following HTML to your site on a relevant page:
<a href="http://absorbance.totallyexplained.com">Absorbance Totally Explained</a>
Then simply click through this link from your web page. Our crawlers will verify your link, extract the title of your web page and instantly add a link back to it. If you like you can remove the words Totally Explained and embed the link in article text.
As long as your link remains in place, we'll keep our link to you right here. Please play fair - our crawlers are watching. Your site must be closely related to this one's topic. Any kind of spamming, dubious practises or removing the link will result in your link from us being dropped and, potentially, your whole site being banned. |